Jaleh B, Asgary A. Precise Water Isotope Analysis in the Natural Concentration Range Using Off-Axis Integrated Cavity Output Spectroscopy (OA-ICOS). IJOP 2024; 18 (1) :43-50
URL:
http://ijop.ir/article-1-575-fa.html
Precise Water Isotope Analysis in the Natural Concentration Range Using Off-Axis Integrated Cavity Output Spectroscopy (OA-ICOS). . 1403; 18 (1) :43-50
URL: http://ijop.ir/article-1-575-fa.html
چکیده: (147 مشاهده)
A compact sensor utilizing off-axis integrated cavity output spectroscopy at a wavelength of 1.389 μm was developed to measure the δ¹⁸O and δ²H isotopic ratios in liquid water samples within the natural concentration range. Allan variance analysis indicated an optimal averaging time of approximately 21 s. The stability and long-term reproducibility of the laser system were assessed using 1-second averaging over 75 min. The minimum detectable absorption (MDA) was determined to be approximately 8.4 × 10⁻⁴ Hz⁻¹/² with a 17 s averaging time. To assess accuracy and reproducibility, the setup was tested within the natural concentration range (δ²H: -400 to 0‰ and δ¹⁸O: -54 to 0‰) using five standard samples, achieving an average accuracy and reproducibility of better than 2.5‰ for δ²H and 0.5‰ for δ¹⁸O.
نوع مطالعه:
پژوهشي |
موضوع مقاله:
اسپکتروسکپی دریافت: 1403/9/28 | ویرایش نهایی: 1404/4/24 | پذیرش: 1404/4/22 | انتشار: 1404/4/24